Microscope specifications for FTD
Written by Michael Krochmal   
Tuesday, 04 September 2007 07:44

In broad terms, what type of microscope do I need for FTD work ?

FTD work is carried out at two magnifications : a low power objective is used to locate good candidate grains, and very high magnification is used to count tracks and measure their length.

Immersion oil is NEVER used, as its refractive index is similar to that of both the mica and the mineral grains, and it will cause any tracks to disappear !

FTD work is carried out in bright-field mode, making use of both transmitted and reflected light. The transmitted light mode results in good images ofcomplete tracks, whereas the reflected light mode is used to accentuate the tracks' etch pit entrances.

In general terms, a microscope should have transmitted and reflected light sources with no less than 20 Watt output, high-quality (high numerical aperture, long working distance and flat field) dry objectives with 10x, 20x, 40x and 100x magnifying powers (a 5x is a possible option, as would be a 50x in place of the 40x), a long-working distance condenser, and high-quality 10x eyepieces.

A so-called trinocular head, which incorporates a dual light path, allowing simultaneous or alternative use of the stereo eye-pieces and a top-mounted camera, is also required. To ensure camera attachment compatibility, the camera port on the microscope must be fitted with a so-called "C-mount" fitting. This is an industry-standard thread fitting which is incorporated in most cameras supplied for this purpose.

One other essential item (but one which many microscopy laboratories may already own) is a high-quality stage micrometer. This is a standard microscope slide bearing a series of very accurately engraved lines at regular and known intervals. The stage micrometer is indispensable for calibration of the various system components.An eyepiece graticule, having a series of numbered and alphabetically indexed squares in a grid arrangement, is also a very useful adjunct. This item is fitted to one of the eyepieces and results in the grid being superimposed over the observed image.

Other items of the system are optional linear external position sensors fitted to the stage for increased stage accuracy, and dual large-format LCD monitors for simultaneous large-image viewing of grain and mica mounts

For our latest software product (the automatic track counting module) to function properly, the following items are essential :
1. a high-quality motorised microscope
2. a high-quality digital camera
3. a very high performance PC
4. An anti-vibration table
5. A high-quality grinding and polishing setup
6. A facility for vacuum deposition of a reflective surface
(These items are explained separately in another FAQ) 

 

Contact Details

as_word.jpg
Autoscan Systems Pty. Ltd.
Street Address : 4/293 Bay Street, Brighton 3186 Victoria Australia
Postal Address : PO Box 112, Ormond 3204 Victoria Australia
Telephone : +61 3 9596 8065 / +61 3 9596 8092
Fax : +61 3 9596 8369
To send us an email - Please click here

Webmaster : Michael Krochmal
This site established 23 October 2007
Supersedes www.autoscan.com.au website established 1995